Abstract
BackgroundInaccurate biometry can lead to the wrong intraocular lens implantation and result in refractive surprise following cataract surgery. It is important to be sceptical of biometry results that do not match the refractive or clinical picture and ask for it to be repeated.Case presentationWe present a unique cause of refractive surprise in a patient undergoing cataract surgery. Pre-operative refraction demonstrated hypermetropia, yet swept-source optical coherence tomography (SS-OCT) biometry repeatedly calculated the axial length as > 35.00 mm in both eyes. The patient underwent phacoemulsification and intraocular lens insertion using the provided biometry calculations, however post-operatively the patient had a + 14.00 dioptre refractive surprise. Analysis of biometry performed on the same day identified other patients with exaggerated axial lengths, supporting the theory that the biometer’s smeared optical surface was responsible. Following servicing of the machine, repeat biometry of the patient calculated the axial length consistent with a hypermetrope (21.67 mm) and the intraocular lens exchange was successful in correcting the refractive error.ConclusionsEnsure the optical surfaces of the biometer are cleaned regularly, and consider repeating biometry on separate days if repeat biometry still is not in keeping with the refractive or clinical picture. Additionally, re-confirm the axial length with another modality.
Highlights
Inaccurate biometry can lead to the wrong intraocular lens implantation and result in refractive surprise following cataract surgery
On analysis of the biometer, the only abnormality detected was a ‘smear’ on its optical surface, which they had removed with a fine brush
swept-source optical coherence tomography (SS-OCT) offers the additional benefit of acquiring measurements even in dense cataracts [9]
Summary
Ensure the optical surfaces of the biometer are cleaned regularly, and consider repeating biometry on separate days if repeat biometry still is not in keeping with the refractive or clinical picture.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.