Abstract
This paper provides a first experimental verification of the use of image-based feedback for refocusing the therapeutic beam in the presence of strongly scattering objects. This capability may be critical in the use of dual mode ultrasound array (DMUA) systems for noninvasive targeting of liver tumors and noninvasive cardiac ablation. In both cases, the target is partially obstructed by the rib cage, which limits the access and distorts the geometrically-focused high-intensity focused ultrasound (HIFU) therapeutic beam. The optimization procedure is based on the use of single-transmit focus (STF) imaging, in which a single transmit imaging beam employing delays derived from phasing for the therapeutic beam is used. We present experimental results of a DMUA refocusing algorithm that selects the control points from the target(s) and the ribs visible in the STF image. In addition, STF images before and after refocusing provide similar feedback by consistently showing increased echogenicity of the target region while the echogenicity of the ribs is not increased (often reduced).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.