Abstract

PurposeTo compare surgical outcomes of SAVI SCOUT reflector localization (SSL) versus wire localization (WL) for breast tumors. MethodsRetrospective review of 42 SSL cases and 42 WL cases. WL patients were consecutively matched for clinical-pathologic features. Final surgical outcome measures were tumor specimen volume, margin status, and re-excision rates. ResultsNo significant differences were present in median specimen volumes (SSL-15.2cm3 vs. WL-16.3cm3), positive margin rate (SSL-9.5% vs. WL-7.1%), close margin rate (SSL-7.1% vs. WL-11.9%) or re-excision rate (SSL-7.1% vs. WL-9.5%). ConclusionSSL is an acceptable alternative to WL with no significant differences in surgical outcomes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.