Abstract

Conventional transmission pulse oximetry is a noninvasive technique for the continuous monitoring of arterial oxygen saturation (SaO2) from peripheral vascular beds such as the finger tip or earlobe. In this paper we propose to exploit the unique transparency of the ocular media to make reflectance pulse oximetry measurements on the retinal fundus. This technique potentially offers significant advantages over conventional pulse oximetry, primarily the ability to monitor cerebral, as opposed to peripheral, oxygen saturation. We have developed an in vitro system to stimulate the retinal circulation and ocular optics. This system consists of a flexible cuvette located in a model eye and an extracorporeal blood circuit to stimulate arterial blood flow. The system was used to investigate the relationship between SaO2 and the R/IR ratio in reflectance pulse oximetry. To enable in vivo measurements to be made, we also modified a standard haptic contact lens to hold the pulse oximeter probe in front of the pupil. In a preliminary study, the lens was fitted to an awake volunteer and cardiac-synchronous signals were detected by the retinal pulse oximeter.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.