Abstract
The forced oscillation technique makes it possible to evaluate the mechanical properties of the respiratory system with a minimum of cooperation. The method is therefore especially useful in children. Impulse oscillometry (IOS) is a commercially available version of this technique. There is, as yet, limited information on reference values for IOS in children. The aim of this study was to extend the reference values for IOS variables and to study their correlation with height, weight and age in healthy children. A sample (n = 360) of children (age 2.1-11.1 years) was measured by using impulse oscillometry (IOS; Jaeger, Würzburg, Germany). The sample was based on children attending kindergarten in Finland and children attending primary school in Sweden. Measurements of respiratory resistance (Rrs) and reactance (Xrs) at 5, 10, 15 and 20 Hz, total respiratory impedance (Zrs) and the resonance frequency (Fr) were made. All variables were related to body height. Most of them were also weakly related to weight. Reference equations for children (height 90-160 cm) are presented.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.