Abstract

The composition of any technical PCB formulation can be determined directly by analyzing the PCB sample by gas chromatography with a flame ionization detector (GC-FID), provided the relative molecular masses of the components are known. The responses of electron capture and selected-ion monitoring, mass-spectra detectors can then be calibrated for individual PCB congeners by correlation of the chromatographic patterns with those of concentrated PCB samples obtained by GC-FID. This procedure, which uses a given technical PCB formulation as a secondary reference standard mixture, is to be preferred over existing calibration methods, when results with ±10% errors are acceptable because commercial PCB formulations cover the whole range of chlorination products.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.