Abstract

Bone densitometry reports a measure of fracture risk in comparison with young adults (T-scores) and age-matched peers (Z-scores). To date, each manufacturer has provided its own reference range resulting in lack of uniformity. The Australia and New Zealand Bone and Mineral Society and Osteoporosis Australia have recognized the need to standardize the reference range and have recommended that data generated by the Geelong Osteoporosis Study (GOS) be used Australia-wide. The GOS recruited a random, population-based sample of adult women and measured bone mineral density (BMD) at the proximal femur and spine using a Lunar DPX-L. These data were used to establish reference ranges for Lunar machines and, using conversion equations, for Norland and Hologic machines. The new standardized Australian reference ranges for BMD will enable consistent diagnosis of osteoporosis and categorization of fracture risk across different types of densitometers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.