Abstract The influence of temperature variations and uncorrect positioning of the cuvette in its holder on the accuracy of absorbance measurements was thoroughly discussed in the case of potassium dichromate and nicotinic acid solutions to be used as reference materials for the periodic control of spectrophotometers. Results showed that temperature control to within ± 0.5 °C is sufficient. On the other hand, the error connected with cell positioning was found to be negligible under the usual working conditions. Finally, the effect of the purity degree on the absorbance values of the above substances was investigated.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.