Abstract

BackgroundThe aim of this study was to calculate 95% reference intervals and double-sided limits of serum alpha-fetoprotein (AFP) and carcinoembryonic antigen (CEA) according to the CLSI EP28-A3 guideline.Material/MethodsSerum AFP and CEA values were measured in samples from 26 000 healthy subjects in the Shuyang area receiving general health checkups. The 95% reference intervals and upper limits were calculated by using MedCalc.ResultsWe provided continuous reference intervals from 20 years old to 90 years old for AFP and CEA. The reference intervals were: AFP, 1.31–7.89 ng/ml (males) and 1.01–7.10 ng/ml (females); CEA, 0.51–4.86 ng/ml (males) and 0.35–3.45ng/ml (females). AFP and CEA were significantly positively correlated with age in both males (r=0.196 and r=0.198) and females (r=0.121 and r=0.197).ConclusionsDifferent races or populations and different detection systems may result in different reference intervals for AFP and CEA. Continuous reference intervals of age changes are more accurate than age groups.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.