Abstract

Reference equivalent threshold sound pressure levels (RETSPLs), were measured at 5 and 6.3 kHz using Telephonics TDH 39 earphones with MX‐41/AR cushions. The purpose of the test was to verify RETSPL values calculated by interpolation from those specified in present International Standards Organization acoustics standards, ISO 389 “Standard reference zero for the calibration of pure‐tone audiometers,” and its amendments. The RETSPL values are intended for the calibration of audiometers using pure tones of fixed frequencies at the preferred frequencies in one‐third octave steps or pure‐tone audiometers having a continuously variable frequency. Mean measured values of RETSPL's at 5 and 6.3 kHz were 10.9 ± 2.6 and 15.1 ± 2.4 dB, in comparison to predicted values which were 13 and 15 dB, respectively.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.