Abstract
This study assessed the reference dose levels for dental panoramic radiography in Gwangju city, South Korea based on the dose width product (DWP) and compared them with those already established elsewhere. A total of 44 panoramic dental radiographic sets (36 digital and 8 analogue panoramic sets) in 41 dental clinics in Gwangju city were chosen. The third quartile DWP was determined from 429 surface dose measurements of the adult surface dose in panoramic dental radiography. The third quartile DWP for panoramic radiography was 60.1 mGy mm. The proposed DWP reference levels of 60.1 mGy mm were less than or equal to those previously reported in other countries, such as Italy and UK, and acceptable for panoramic radiography in Gwangju, South Korea.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.