Abstract

The neutron source measurement service was established in 1960, based on absolute measurements in a flow-type manganese sulphate bath system, the hydrogen/manganese cross-section ratio having been determined in a dilution experiment. As a result of improvements in chemical analysis of trace elements in the bath solution and in computerized data fitting techniques over this 20-year period, it is possible to re-evaluate the emission rates of neutron sources which have been measured during this period. The re-evaluation demonstrates that the emission rates should be increased by 0.5% mainly due to neutron absorption in traces of boron, cadmium and lithium. The manganese sulphate solution has now been replaced by purer material, and the re-measurement of the cross-section ratio is in progress.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.