Abstract

In a retrospective analysis of vascular interventional procedures, relations between parameters of the examination and radiation exposure of patient and medical personnel are examined. 1208 vascular interventional procedures are evaluated. Interventional procedures are divided into three groups: percutaneous transluminal angioplasty, implantation of a stent, thrombolysis. Mean values of the radiation dose of patient and radiology personnel are reported for these examinations. The mean value of the radiation dose of the physician was 7 microSv (maximum 24 microSv), that of the patient 1548 cGy.cm2 (maximum 8485 cGy.cm2). The quantity of X-rays to the patient may be lowered by using pulsed fluoroscopy and by reducing the number of radiographs. Reduction of the number of radiographs may be achieved by using the last-image hold and the road mapping mode. The operator's dose can be decreased by using additional radiation protection systems like a MAVIG-radiation protection wall. The radiation dose reduction was 61% for the physician and 17% for the patient.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.