Abstract

This paper presents the latest implementation of the ESTI reference cell set at the European solar test installation. The set consists of five photovoltaic reference cells used to measure solar irradiance. Each cell has been calibrated multiple times by primary methods. The stability of the set has been maintained and monitored over a long time period (up to 20 years). Taken together this allowed a rigorous determination of the assigned calibration value for each cell based on the weighted average, together with its uncertainty. This has led to unprecedented low uncertainties of 0.23% (k = 2), which are at least a factor of two smaller than achievable with any single primary calibration. Using these cells for calibrating secondary reference cells with a dedicated set-up leads to uncertainties for these secondary calibrated cells which is of the same order as the best primary calibration methods.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.