Abstract
The triple reamer is a convenient lool for reaming the proximal femur when using a sliding screw and plate device, but unnecessarily high temperatures may be generated within the femoral head while the lateral cortex is being reamed. Nineteen female patients over 60 years old were randomized to either the standard or a modified reaming technique with the triple reamer during internal fixation of proximal femoral fractures. Temperatures generated within the femoral head were measured. Results showed a mean peak temperature of 51.9° C (range 42–66) using the standard method and 46.3° C (range 40–52) using the modified method (P = 0.07). A significant difference (P < 0.05) was found in the duration of temperature elevation above the critical 44° C (17.4s vs 5.7s), it being prolonged in the standard group. Even in this osteoporotic patient group, damaging temperatures which may contribute to the mechanical failure of fixation, are generated within the femoral head. A simple method to avoid this problem is described.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.