Abstract

The primary goal of cancer surgery is to completely eliminate tumors. A real-time diagnostic method uses a fluorescence contrast agent and a surgical microscope to assess the status of tumor resection and the patient’s blood circulation. The biggest problem in imaging diagnostics using a microscope is the specular reflection phenomenon. While observing a lesion, the observation field may be obstructed due to specular reflection, making it difficult to obtain accurate results during the diagnostic process. Herein we propose a method to reduce specular reflection during tumor diagnosis by introducing a linearly polarized filter for a surgical microscope system. The method of angular direction adjustment of the filter ensures that only the horizontally polarized light passes through it, thereby obstructing the specular reflection. As a result of removing specular reflection, clear images were obtained at 90° and 270°. This experiment was conducted using phantoms and animals. Our results prove that the proposed method can be applied to imaging cameras used in internal medicine, surgery, and radiology for diagnosis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.