Abstract

To measure the effect of field of view (FOV) size and position on scatter-induced image noise in cone beam computed tomography (CBCT). A polymethyl methacrylate (PMMA) phantom containing air and aluminum underwent CBCT scanning, using seven FOVs ranging between 4 × 4 cm and 14 × 10 cm, positioned both centrally and according to a dental scan. Signal difference to noise ratio (SDNR) was measured on two-dimensional (2-D) projection images. At a central position, SDNR decreased with increasing FOV size, ranging between 9.8 (14 × 10 cm) and 10.9 (4 × 4 cm). For dental FOV positions, SDNR values were between 6.3 (14 × 10 cm) and 9.5 (4 × 4 cm). To reach a constant SDNR, a dose reduction up to 76% was possible for small FOVs compared with the 14 × 10 cm FOV. The use of small FOVs and peripheral FOV positioning decreases scatter at the detector, resulting in a considerable potential for reduction of radiation dose to the patient.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call