Abstract

In an attempt to reduce the radiation dose during the examination of scoliotic patients, several screen-film combinations have been compared with a conventional system used at present. Kodak's Lanex Regular screen with Kodak Ortho H film enables the dose to be reduced eight times without significant deterioration of the image quality. The dose to the mammary glands can be reduced further by a factor of five if posterior--anterior instead of anterior--posterior projection is used.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.