Abstract

Sound pressure levels were measured within an MR scan room for a range of sequences employed in veterinary brain scanning, using a test phantom in an extremity coil. Variation of TR and TE, and use of a quieter gradient mode (‘whisper’ mode) were evaluated to determine their effect on sound pressure levels (SPLs). Use of a human head coil and a human brain sequence was also evaluated. Significant differences in SPL were achieved for T2, T1, T2* gradient echo and VIBE sequences by varying TR or TE, or by selecting the ‘whisper’ gradient mode. An appreciable reduction was achieved for the FLAIR sequence. Noise levels were not affected when a head coil was used in place of an extremity coil. Due to sequence parameters employed, veterinary patients and anaesthetists may be exposed to higher sound levels than those experienced in human MR examinations. The techniques described are particularly valuable in small animal MR scanning where ear protection is not routinely provided for the patient.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.