Abstract
The present paper outlines the methods used to reduce the interference signal caused by ascorbic and uric acids on an ammonia amperometric sensor in aqueous solutions. The sensor was prepared by electrochemical deposition of poly(pyrrole), doped with dodecylbenzenesulfonic anions (DBS), onto platinum substrates. Poly(pyrrole) films were coated with different membranes, aiming to find different ways to diminish the interference response of anionic species such as ascorbate and urate. Nafion ®, cellulose acetate and poly(5-amino-1-naphthol), were studied as permselective layers. This last polymer is electroactive, offering the possibility of its formation by electrochemical deposition onto poly(pyrrole). In order to verify the stability of poly(5-amino-1-naphthol) when used in basic medium, Raman spectra were taken from these modified electrodes. Further, all sensors showed a linear response and a good sensitivity in a concentration range from 0 to 120 μmol L −1 which covers the ammonia concentration range typical of clinical samples (18–72 μmol L −1).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.