Abstract

ObjectivesTo evaluate the reduction of artifacts from cardiac implantable electronic devices (CIEDs) by virtual monoenergetic images (VMI), metal artifact reduction (MAR) algorithms, and their combination (VMIMAR) derived from spectral detector CT (SDCT) of the chest compared to conventional CT images (CI).MethodsIn this retrospective study, we included 34 patients (mean age 74.6 ± 8.6 years), who underwent a SDCT of the chest and had a CIED in place. CI, MAR, VMI, and VMIMAR (10 keV increment, range: 100–200 keV) were reconstructed. Mean and standard deviation of attenuation (HU) among hypo- and hyperdense artifacts adjacent to CIED generator and leads were determined using ROIs. Two radiologists qualitatively evaluated artifact reduction and diagnostic assessment of adjacent tissue.ResultsCompared to CI, MAR and VMIMAR ≥ 100 keV significantly increased attenuation in hypodense and significantly decreased attenuation in hyperdense artifacts at CIED generator and leads (p < 0.05). VMI ≥ 100 keV alone only significantly decreased hyperdense artifacts at the generator (p < 0.05). Qualitatively, VMI ≥ 100 keV, MAR, and VMIMAR ≥ 100 keV provided significant reduction of hyper- and hypodense artifacts resulting from the generator and improved diagnostic assessment of surrounding structures (p < 0.05). Diagnostic assessment of structures adjoining to the leads was only improved by MAR and VMIMAR 100 keV (p < 0.05), whereas keV values ≥ 140 with and without MAR significantly worsened diagnostic assessment (p < 0.05).ConclusionsThe combination of VMI and MAR as well as MAR as a standalone approach provides effective reduction of artifacts from CIEDs. Still, higher keV values should be applied with caution due to a loss of soft tissue and vessel contrast along the leads.Key Points• The combination of VMI and MAR as well as MAR as a standalone approach enables effective reduction of artifacts from CIEDs.• Higher keV values of both VMI and VMIMARat CIED leads should be applied with caution since diagnostic assessment can be hampered by a loss of soft tissue and vessel contrast.• Recommended keV values for CIED generators are between 140 and 200 keV and for leads around 100 keV.

Highlights

  • Metal artifacts arise as a combination of beam-hardening which results from absorption of low energetic photons [6, 7], photon starvation which is caused by an insufficient amount of photons reaching the detector [7, 8], and scatter artifacts [9]

  • We considered the standard deviation (SD) in artifact-impaired tissue indicative for artifact burden [18], it needs to be considered that SD depends on several factors that might bias artifact reduction and changes in SD can be rather subtle even though changes in mean attenuation can be large

  • The manufacturers and models of the cardiac implantable electronic devices (CIEDs) generator and leads are provided in Table 1 of the supplementary material

Read more

Summary

Objectives

To evaluate the reduction of artifacts from cardiac implantable electronic devices (CIEDs) by virtual monoenergetic images (VMI), metal artifact reduction (MAR) algorithms, and their combination (VMIMAR) derived from spectral detector CT (SDCT) of the chest compared to conventional CT images (CI). Mean and standard deviation of attenuation (HU) among hypo- and hyperdense artifacts adjacent to CIED generator and leads were determined using ROIs. Two radiologists qualitatively evaluated artifact reduction and diagnostic assessment of adjacent tissue. Results Compared to CI, MAR and VMIMAR ≥ 100 keV significantly increased attenuation in hypodense and significantly decreased attenuation in hyperdense artifacts at CIED generator and leads (p < 0.05). VMI ≥ 100 keV, MAR, and VMIMAR ≥ 100 keV provided significant reduction of hyper- and hypodense artifacts resulting from the generator and improved diagnostic assessment of surrounding structures (p < 0.05). Conclusions The combination of VMI and MAR as well as MAR as a standalone approach provides effective reduction of artifacts from CIEDs. Still, higher keV values should be applied with caution due to a loss of soft tissue and vessel contrast along the leads. Key Points The combination of VMI and MAR as well as MAR as a standalone approach enables effective reduction of artifacts from

Methods
Results
Discussion
Conclusion