Abstract

To solve the problem of carry-over – a persistent chromatographic challenge for bioanalytical assays with highly sensitive detector such as a mass spectrometry (MS) – a new on-line sample pretreatment HPLC/MS system using a column-switching technique was established. This system was designed to reduce carry-over based on a hydrophobic interaction mechanism using a washing function (multi-mobile phase flow system), as well as to remove impurities on-line in a mobile phase for the pretreatment. As a result, a washing function in this system was enabled to reduce carry-over and to remove impurities in a mobile phase by automatic operation. Therefore, concentration levels of donepezil (DH) in dog plasma as low as 10 pg/mL could be determined using this on-line sample pretreatment HPLC/MS system. In addition, method validation results of specificity, linearity, accuracy, precision, lower limit of quantitation (LLOQ), and carry-over demonstrated that this on-line sample pretreatment HPLC/MS system was robust and valid as a practical assay of drugs and metabolites in the biological samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.