Abstract

Objective To assess the effectiveness of Z-axis automatic tube current modulation ( ATCM) technique with respect to dose reduction and image noise in the thyroid regional tissues during neck computed tomography angiography (CTA). Methods A total of 140 patients underwent neck CTA with 64-slice spiral CT, and were randomized into fixed-tube current group and ATCM group. Objective noise, CTDIw, DLP and mA of each group were recorded. Results The image qualities of all the patients were diagnostically acceptable, though the objective noise of neck in the ATCM group was higher than that in the group of fixed-tube current (P< 0.05 ). The radiation doses in the ATCM group [ CTDIw (35.99±1.31 ) mGy, DLP (11 121.39 ±5.51) mGy·cm] were significantly inferior to those in the group of fixed tube current [ CTDIw (43.22±1.42) mGy, DLP (1514.45±5.56) mGy·cm]. Conclusions Compared with fixedtube current technique, ATCM technique could significantly reduce the radiation dose in neck CT with diagnostic acceptability of the image quality. Key words: Computed tomography angiography (CTA); Automatic tube current modulation ( ATCM ); Radiation protection

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.