Abstract

The purpose of this study was to determine whether an alternate fitting strategy, specifically adjustment to gains in a hearing aid (HA), would improve performance in patients who experienced poorer performance in the bimodal condition when the HA was fit to traditional targets. This study was a retrospective chart review from a local clinic population seen during a 6-month period. Participants included 6 users of bimodal stimulation. Two performed poorer in the cochlear implant (CI) + HA condition than in the CI-only condition. One individual performed higher in the bimodal condition, but the overall performance was low. Three age range-matched users whose performance increased when the HA was used in conjunction with a CI were also included. The HA gain was reduced beyond 2000 Hz. Speech perception scores were obtained pre- and postmodification to the HA fitting. All listeners whose HA was programmed using the modified approach demonstrated improved speech perception scores with the modified HA fit in the bimodal condition when compared with the traditional HA fit in the bimodal condition. Modifications to gains above 2000 Hz in the HA may improve performance for bimodal listeners who perform more poorly in the bimodal condition when the HA is fit to traditional targets.