Abstract
Overlapping biphasic (OLBI) stimulation is a new pacing waveform consisting of two simultaneous monophasic pulses of opposite polarities applied to a bipolar electrode. The goal of this prospective study was to compare, using conventional pacing leads, the acute energy pacing thresholds, measured at 0.5-ms pulse duration, associated with bipolar versus OLBI (mode 7 and mode 8) pulse delivery. Thirty one leads were tested in 20 patients. Of these leads, 7 (23%) were implanted chronically, 12 (39%) were in atrial positions, and 19 (61%) in ventricular positions. Energy pacing thresholds were significantly lower (-25.6 +/- 25.6%, P = 0.005) in OLBI mode 8 (1.30 +/- 3.96 microJ) compared to bipolar (1.55 +/- 4.37 microJ) pacing, regardless of the pacing site or length of service of the leads. In contrast, OLBI mode 7, which has an anodal component, at the tip electrode was associated with higher energy pacing thresholds (3.65 +/- 6.48 microJ; +358.3 +/- 219.4%, P = 0.002). Mode 8 OLBI pacing is associated with lower acute energy pacing thresholds when used with bipolar leads in contact with the myocardium. OLBI pacing may increase pacemaker longevity by reducing long-term energy consumption.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.