Abstract

Excessive tissue-instrument interaction forces during robotic surgery have the potential for causing iatrogenic tissue damages. The current in vivo study seeks to assess whether tactile feedback could reduce intraoperative tissue-instrument interaction forces during robotic-assisted total mesorectal excision. Five subjects, including three experts and two novices, used the da Vinci robot to perform total mesorectum excision in four pigs. The grip force in the left arm, used for retraction, and the pushing force in the right arm, used for blunt pelvic dissection around the rectum, were recorded. Tissue-instrument interaction forces were compared between trials done with and without tactile feedback. The mean force exerted on the tissue was consistently higher in the retracting arm than the dissecting arm (3.72 ± 1.19 vs 0.32 ± 0.36N, p < 0.01). Tactile feedback brought about significant reductions in average retraction forces (3.69 ± 1.08N vs 4.16 ± 1.12N, p = 0.02), but dissection forces appeared unaffected (0.43 ± 0.42 vs 0.37 ± 0.28N, p = 0.71). No significant differences were found between retraction and dissection forces exerted by novice and expert robotic surgeons. This in vivo animal study demonstrated the efficacy of tactile feedback in reducing retraction forces during total mesorectal excision. Further research is required to quantify the clinical impact of such force reduction.