Abstract
A method is presented in which the couch position is predicted before the treatment instead of obtaining a reference position at the first treatment fraction. This prevents systematic differences in patient posture between preparation and treatment. In literature, only limited data are available on couch positioning. We position our patients at the planned couch position, allowing a small difference between skin marks and lasers, followed by online imaging. For a 3-month period, our standard deviations (mm) in couch position in the vertical, longitudinal, and lateral directions were head and neck—1.6, 2.8, and 2.5; thorax—2.9, 5.5, and 4.5; breast—3.0, 4.1, and 4.0; and pelvis—3.5, 4.0, and 4.7, respectively. We have improved the reproducibility of patient posture in our institute by using the predicted couch position. Our data may serve as a reference for other institutes because the couch position variation is less than that published in literature.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.