Abstract
For an ion trap with resonance ejection, peak intensity and peak position of the acquired mass spectra are affected by the phase difference between the radio frequency (RF) and auxiliary alternating current (AC) potentials. To ensure measurement stability, RF and AC phase-locking is commonly used in commercial ion trap mass spectrometers. In this study, a compact electronic control system was developed to accurately regulate the RF and AC phases and was employed in a photoionization rectilinear ion trap (RIT) mass spectrometer. We found that the phase-locking method was defective in multicomponent analysis because the optimal RF and AC phase difference was usually different for different m/z peaks. After studying and characterizing the relationship between the peaks and the RF and AC phases, a correction method based on data processing was used to improve the peaks' stability and accuracy. The results show that the fluctuations of both peak intensity and peak position were significantly reduced and that the instrument presented satisfying reproducibility and quantitative ability.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.