Abstract

This study developed algorithms to decrease the arrhythmia false alarms in the ICU by processing multimodal signals of photoplethysmography (PPG), arterial blood pressure (ABP), and two ECG signals. The goal was to detect the five critical arrhythmias comprising asystole (ASY), extreme bradycardia (EBR), extreme tachycardia (ETC), ventricular tachycardia (VTA), and ventricular flutter or fibrillation (VFB). The different characteristics of the arrhythmias suggested the application of individual signal processing for each alarm and the combination of the algorithms to enhance false alarm detection. Thus, different features and signal processing techniques were used for each arrhythmia type. The ECG signals were first processed to reduce the signal interference. Then, a Hilbert-transform based QRS detector algorithm was utilized to identify the QRS complexes, which were then processed to determine the instantaneous heart rate. The pulsatile signals (PPG and ABP) were processed to discover the pulse onset of beats which were then employed to measure the heart rate. The signal quality index (SQI) of the signals was implemented to verify the integrity of the heart rate information. The overall score obtained by our algorithms in the 2015 Computing in Cardiology Challenge was a score of 74.03% for retrospective and 69.92% for real-time analysis.