Abstract
In color flow imaging for medical diagnosis, the inherent trade-off between frame rate and image quality may often lead to suboptimal images. Parallel receive beamforming is used to help overcome this problem, but this introduces artifacts in the images. In addition to the parallel beamforming artifacts found in B-mode imaging, we have found that a difference in curvature of transmit and receive beams gives a bias in the Doppler velocity estimates. This bias causes a discontinuity in the velocity estimates in color flow images. In this work, we have shown that interpolation of the autocorrelation estimates obtained from overlapping receive beams can reduce these artifacts significantly. Because the autocorrelation function varies quite slowly, the beams can be acquired with a considerable time difference, for instance across interleaving groups or across scan planes in a 3-D scan. We have shown that a high frame rate of color flow images can be maintained with parallel beam acquisition with minimal deterioration of the image quality.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.