Abstract

Magnesia-stabilized zirconia (Mg-PSZ) is stable and maintains a scratch-resistant surface in hip replacement, but is untested in knees. We assessed whether using Mg-PSZ instead of cobalt-chromium (CoCr) femoral components resulted in less tibial insert wear, and evaluated changes in topography and roughness of the femoral components. Inserts bearing against CoCr or Mg-PSZ were tested using standard (9Mc) and aggressive (6Mc) waveforms. Femoral component surface topography and roughness were evaluated before and after testing by optical profilometry. When bearing against Mg-PSZ, UHMWPE wear rate decreased by 73% (standard) and by 59% (aggressive conditions). After 15Mc, CoCr components featured deep scratches, and roughness increased five-fold, while Mg-PSZ components were unchanged. Mg-PSZ femoral components may be indicated for high-demand patients and those with metal sensitivity.