Abstract
The goal of this investigation was to examine the influence of two oxygenators with different membranes, made of either polypropylene (PPL) or polymethylpentane (PMP), on the plasma concentration of sevoflurane during cardiopulmonary bypass. The concentrations of sevoflurane during cardiopulmonary bypass were examined in patient plasma, endotracheal tubes, cardiotomy reservoirs and the outlets of the heart-lung oxygenators in twenty patients who underwent elective heart surgery. The sevoflurane losses are smaller in cardiopulmonary bypass when using a polymethylpentane versus a polypropylene oxygenator. Ten minutes after beginning cardiopulmonary bypass, the sevoflurane plasma concentration in the PPL oxygenator group compared to the PMP oxygenator group fell significantly (PPL 0.48-1.79 (0.93) vs. PMP 0.80-2.15 (1.56) µL × 100 mL-1, p = 0.02). This difference persisted until ten minutes after the termination of cardiopulmonary bypass. The results of this study show that using a polymethylpentane membrane oxygenator rather than a polypropylene oxygenator significantly reduces the losses of sevoflurane, resulting in higher plasma concentrations and greater depth of anesthesia.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.