Abstract

During total knee replacement, hinged total knee implants are used in cases where ligament balancing cannot be achieved with less-constrained implants. The case of a patient who experienced two episodes of intraprosthetic dislocation of his rotating-hinge total knee prosthesis is described. There are very few reports of this type of dislocation with these implants. The implant's design, particularly of the hinge, plays an important role in stability. The balance between the flexion and extension spaces is very important even when using a hinged total knee implant. The role of the extensor mechanism in anteroposterior stability is reviewed, along with simple ways to augment it.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.