Abstract

Gadolinium aluminum gallium garnet (GAGG) is a promising scintillator crystal for positron emission tomography (PET) detectors owing to its advantages of energy resolution, light yield, and absence of intrinsic radiation. However, a large portion of the incident photons undergoes Compton scattering within GAGG crystal because of its low stopping power compared to that of lutetium-based crystals such as Lu2SiO5 (LSO). Inter-detector scattering (IDS) and inter-crystal scattering (ICS) result in loss of sensitivity and image quality of PET, respectively. We performed a Monte Carlo simulation study to evaluate IDS recovery in our currently developing brain-dedicated PET, and extended the idea to ICS recovery. We also compared the impact of the recoveries on LSO- and GAGG-based PET scanners. We measured the sensitivity and spatial resolution of the brain PET, and analyzed the image quality using a lesion phantom, a hot-rod phantom, and a 2D Hoffman phantom with applying IDS or ICS recovery. IDS recovery increased the PET sensitivity and improved the noise level of the reconstructed images. ICS recovery enhanced the spatial resolution and the contrast of the images was improved. As the occurrence rates of IDS and ICS were higher in GAGG than in LSO, the overall impact of IDS or ICS recovery was significant in GAGG. In conclusion, we showed that the proportional method would be suitable for IDS and ICS recoveries of PET, and emphasized the importance of ICS and IDS recoveries for PET using crystals with low stopping power.