Abstract
We report the recovery of Coccidioides immitis from the blood and abscess fluid of two separate patients by using two automated blood culture systems. In the first case, an aspirate from a neck abscess containing C. immitis spherules was serially diluted and inoculated into liquid media used by the BacT/Alert and the Bactec NR660 blood culture systems. BacT/Alert bottles inoculated with 10 5, 10 4, 10 3, 10 2, 10, and two spherules produced a positive signal at 19, 24, 35, 42, 57, and 62 h postinoculation, respectively. Bactec NR660 bottles containing >10 2 spherules and 10 spherules produced a positive signal after approximately 72 and 96 h of incubation, respectively. In the second case, a blood specimen incubated in BacT/Alert blood culture broth was signaled positive after 82 h of incubation. No organisms were detected by Gram stain of the broth, but C. immitis grew after blind subculture. Our observations demonstrate that these rapid blood culture systems are capable of supporting growth of C. immitis. To our knowledge, this report is the first to detect C. immitis by these blood culture systems.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.