Abstract

A retrospective study of 68 patients was carried out to assess the complications arising from the use of reconstruction plates in the maintainance of space and contour following mandibular segmental resection. Skin or mucosal perforation, plate fracture and loss of screw retention were the main complications. The most susceptible sites to screw loosening in the plates were situated nearest to and farthest from the resection margin on the proximal residual segments. Some of the possible causative biomechanical mechanisms of plate failure were studied using photoelastic models simulating the types of mandibular defect and plate fixation. Isochromatic fringe analysis was used to analyse stress lines in the bone surrounding serews. It was found that, during functional loading, moment and shear forces produced high concentrations of squeeze and press stress in this bone, causing bone resorption with consequent loss of screw retention.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.