Abstract

A greater flexion angle in total knee arthroplasty (TKA) is desirable for replacing the human knee joint damage of people living in the Middle East and Asian regions. This flexion angle is a significant clinical factor affecting the range of motion for patients. Therefore, this study aims to optimise the flexion angle in the design of the Artificial Knee Joint (AKJ) by tailoring the Posterior tibial slope (PTS) and thickness of the Posterior Femoral Condyle (PFC). The 3D CAD model of AKJ was obtained from the reverse engineering (RE) process. Furthermore, the physical model used as a standard implant produced by DePuy, Cruciate Substituting (CS)-Revision type. The obtained 3D models were subsequently converted into CAD for the reconstruction process. Reconstruction of the CAD Model with the customised PTS and FPC components provided the flexion angle in the range of 149.9°–166.7°, which is required for the motion suitable for Middle Eastern and Asian people. The results may be used as a reference standard for doctors in hospitals or industries to design AKJ with Asian ergonomics.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.