Abstract

BackgroundThe convectional strain-based algorithm has been widely utilized in clinical practice. It can only provide the information of relative information of tissue stiffness. However, the exact information of tissue stiffness should be valuable for clinical diagnosis and treatment.MethodsIn this study we propose a reconstruction strategy to recover the mechanical properties of the tissue. After the discrepancies between the biomechanical model and data are modeled as the process noise, and the biomechanical model constraint is transformed into a state space representation the reconstruction of elasticity can be accomplished through one filtering identification process, which is to recursively estimate the material properties and kinematic functions from ultrasound data according to the minimum mean square error (MMSE) criteria. In the implementation of this model-based algorithm, the linear isotropic elasticity is adopted as the biomechanical constraint. The estimation of kinematic functions (i.e., the full displacement and velocity field), and the distribution of Young’s modulus are computed simultaneously through an extended Kalman filter (EKF).ResultsIn the following experiments the accuracy and robustness of this filtering framework is first evaluated on synthetic data in controlled conditions, and the performance of this framework is then evaluated in the real data collected from elastography phantom and patients using the ultrasound system. Quantitative analysis verifies that strain fields estimated by our filtering strategy are more closer to the ground truth. The distribution of Young’s modulus is also well estimated. Further, the effects of measurement noise and process noise have been investigated as well.ConclusionsThe advantage of this model-based algorithm over the conventional strain-based algorithm is its potential of providing the distribution of elasticity under a proper biomechanical model constraint. We address the model-data discrepancy and measurement noise by introducing process noise and measurement noise in our framework, and then the absolute values of Young’s modulus are estimated through the EFK in the MMSE sense. However, the initial conditions, and the mesh strategy will affect the performance, i.e., the convergence rate, and computational cost, etc.

Highlights

  • In the daily clinical exercises, palpation is frequently used as one part of physical examination to determine suspicious lesion’s size, shape, firmness, or location [1], but this process is largely decided by the doctor’s experiences

  • The problem could be tackled in three dimension to avoid the throughplane motion effect, but in this paper we only study the two-dimensional problem because the images of quasi-static ultrasound elastography are two dimensional data collected by the linear array of ultrasonic probe

  • Since there are not much previous works to study the properties of process noise, the process noise is modeled as white noise distribution, which is the contribution of this paper for iterative approaches to quasi-static elastography to our knowledge

Read more

Summary

Introduction

In the daily clinical exercises, palpation is frequently used as one part of physical examination to determine suspicious lesion’s size, shape, firmness, or location [1], but this process is largely decided by the doctor’s experiences. After the deformation caused by the external loading is captured by the ultrasound machine, the elastographic image can be generated using different reconstruction techniques [22]. After the shear wave of the biological tissue is captured by the ultrasound machine, the following images of shear modulus can be generated using similar reconstruction techniques [24]. The convectional strain-based algorithm has been widely utilized in clinical practice. It can only provide the information of relative information of tissue stiffness. The exact information of tissue stiffness should be valuable for clinical diagnosis and treatment

Methods
Results
Discussion
Conclusion