Abstract
In a clinical contest, it is common to use dedicated phantoms to perform quality assurance test to check the performance of a SPECT system. Some of these phantoms are also used to calibrate the system for dosimetric evaluation of patients undergoing radiometabolic cancer therapy. In this work, a 3D-OSEM reconstructed 177Lu SPECT dataset of a homogeneous cylindrical phantom is described. This dataset was acquired to investigate the variation of the SPECT calibration factor, counts convergence, noise and uniformity by varying the number of subsets and iterations. In particular, the dataset is composed of images reconstructed using five different numbers of subsets and sixteen different numbers of iterations, for a total of 80 different configurations. The dataset is suitable for comparison with other reconstruction algorithms (e.g. FBP, MLEM, etc.) and radionuclides (e.g. technetium, yttrium). In regards to the uniformity issue, the same dataset allows the user to perform radiomic investigations on the influence of the border effect on the reconstructed images.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.