Abstract
We present a System-On-Chip Field Programmable Analog Array (FPAA) for analyzing and processing the signals off an accelerometer for a wearable joint health assessment device. FPAAs have been shown to compute with an efficiency of 1000 times, as well as area efficiencies of 100 times, more than digital solutions. This work presents a low power signal processing system which allows us to extract features from the output of the accelerometer. These features are used by the classifier, implemented using a vector matrix multiplication and a two output 1-winner-take-all, to detect flexion and extension cycles in the subject. The compiled design consumes 0.636 μW of power for the front end analog signal processing chain where as the single layer classifier uses 13 μW of power. Thus the system is highly suitable for wearable applications where power consumption is a major concern. The current FPAA is fabricated in a 0.35 μm CMOS process and is operated at a power supply of 2.5 volts. The Gm-C filters and other circuits are operated in the subthreshold regime of the transistor to obtain the highest transconductance to current ratio offered by the process.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.