Abstract

This work reports a comparison of an enzyme immunoassay (EIA) using two major Treponema pallidum recombinant antigens with a T. pallidum hemagglutination (TPHA) assay and a nontreponemal Venereal Disease Reference Laboratory (VDRL) test. A total of 1,822 normal donor serum samples was tested for cardiolipin and T. pallidum antibodies, respectively, by the VDRL assay and EIA. Among these samples, 440 were further tested by TPHA technology. Four samples were found positive by EIA, while all were reported to be negative by both TPHA and VDRL routine assays. Subsequent testing of EIA-positive samples confirmed 100% (four of four samples) and 25% (one of four samples) positive results, respectively, by immunofluorescence assay and a Western blot (immunoblot) syphilis kit. The sensitivity of the recombinant EIA was estimated at virtually 100% with a reference panel of 50 syphilitic samples. According to this study, the newly developed EIA kit shows 100% sensitivity combined to a specificity greater than 99.8% for detecting treponemal immunoglobulin G antibodies in blood bank syphilis screening.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.