Abstract

Microphones can be calibrated by a primary technique in a diffuse sound field. The formula for the diffuse-field voltage response is derived from the well-known relationship for a reciprocal transducer in free space. It is shown that the reciprocity parameter for a diffuse-sound field follows from that for free space by replacing the distance by the “diffuse-field distance” of a point source. This distance depends only on the total absorption in the reverberation room. The experimental procedure is described in some detail and the results of the diffuse-field calibration of a Western Electric 640 AA condenser microphone are presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.