Abstract
We report the recent advance in our rubidium atomic fountain clock (AFC). The parameters of the Ramsey cavity are optimized by balancing the coupling from the two ports. The temperature control system of the Ramsey interaction region is renovated, and the resonant temperature of the Ramsey cavity is regulated to be slightly above the room temperature. The quality of magnetic environment in the Ramsey interaction region is also improved. A new digital-to-analog converter (DAC) circuit that controls the local oscillator is adopted to decrease the noise of the oven-controlled crystal oscillator output. As a result, the short-term fractional frequency stability of 2.7 × 10−13τ−1/2 and the long-term fractional frequency stability of 1.6 × 10−15 at the average time of 32800 s are achieved.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.