Abstract

Photothermal Deflection Spectroscopy (PDS) is a recently developed technique that is finding a useful application in the measurement of low optical absorptance of thin films. Among the noise sources affecting the PDS measurement, probe beam pointing instability and mechanical vibration play a considerable role. In this work an optoelectronic system for the reduction of their influence is described. Moreover, PDS measurements are typically performed keeping the sample immersed in a deflecting liquid; thus measured values of absorptance must be corrected when other surrounding media, as air, are considered. This correction is an easy task for single film coatings. Here the general case of an unknown multiplayer coating is analysed; a range of values containing the true absorptance in air is obtained by theoretical analysis and a practical method to evaluate the absorptance in air is discussed. Finally, deflecting liquids alternative to the commonly used CCI4 have been examined. Useful optical range, thermal diffusivity and “relative deflecting power” of CCI4, CS2, Iso-octane and Aceton are reported.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.