Abstract
An experimental setup used at the CENTA (Centre for Nuclear and Accelerator Technologies) laboratory to carry out IBA (Ion Beam Analysis) techniques is described. PIXE (Particle Induced X-ray Emission) technique offers a unique way of non-destructive elemental analysis using accelerated ion beams. The calibration of instrumental constants (H-values) of SDD and BEGe detectors, used in the setup, is described in detail. The H-value was determined as a function of X-ray energy by using thin MicroMatter standards of chosen elements. The calibrated H-values were then used in GUPIXWIN for the determination of elemental concentrations in two IAEA reference materials, which were compared with reference values. A reasonable agreement was achieved between the measured and reference values, while the calibrated H-values will be used for further analyses.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.