Abstract

The COMET experimental station at SEXTANTS beamline of the SOLEIL synchrotron is dedicated to the coherent imaging of magnetic domains, using holography and ptychography, in a sample environment including magnetic fields up to 0.9 T, temperature range from 30 to 400 K and RF pumping. The 2D-detector capabilities of COMET were recently upgraded by adding to the standard CCD camera two new detectors: a high repetition rate 2D CMOS and a double delay line MCP for time resolved experiments. The three detectors can be interchanged within minutes without breaking the vacuum. The distance between an individual detector and the sample can be varied over a wide range, allowing for a trade-off between pixel size and field of view in the image reconstruction. A reconstructed pixel size of the order of 10 nm has been achieved with the new COMET setup. We report also a comparison between holographic images of magnetic domains obtained for the same sample by using the CMOS and the CCD detectors

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.