Abstract

Short-wavelength synchrotron radiation excitation has been an indispensable tool in the studies of the properties of wide gap materials using time-resolved low-temperature luminescence spectroscopy. In recent years, several setups for such investigations have been launched at MAX IV Laboratory and Photon Science at DESY. Two permanently stationed time-resolved luminescence setups at FinEstBeAMS and P66 beamlines are in operation at MAX IV 1.5 GeV and Petra III storage rings, respectively. Mobile luminescence setups have been developed for studies at FemtoMAX and P23 beamlines. FinEstBeAMS, P66 and P23 provide time resolution from ∼160 to 100 ps. The FemtoMAX photon source based on an in-vacuum undulator getting an electron beam from the 3 GeV linear accelerator provides an exceptional time resolution of ∼30 ps, limited by time response of the photodetector. The performance of the setups, achieved milestones and research challenges are discussed for four new luminescence stations available for the research community with the main focus on time-resolved techniques.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call