Abstract
A limiting factor of tuneable diode laser spectroscopy (TDLS) with wavelength modulation spectroscopy (WMS) is the presence of background residual amplitude modulation (RAM) on the recovered 1<sup>st</sup> harmonic signal. The presence of this background term is due to direct modulation of the source laser power. This work presents a novel method to optically remove the unwanted background, with the major benefit being that measurement sensitivity can be increased. The recently developed phasor decomposition method1 (PDM), is a near IR (NIR) TDLS analysis technique that is used with the addition of the new RAM nulling method to recover gas absorption line-shapes. The PDM is a calibration free approach, which recovers the gas absorption line-shape and the isolated 1<sup>st</sup> derivative of the line-shape from the 1<sup>st</sup> harmonic signal. The work presented illustrates and validates the new RAM nulling procedure with measurements examining the 1650.96nm absorption line of methane (CH<sub>4</sub>) with comparisons to theory.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.