Abstract

We describe the recent progress on the use of Schottky Cadmium telluride (CdTe) diode detectors for γ-ray spectroscopy. The extremely low leakage current of the newly developed CdTe diode allows us to apply a much higher bias voltage in comparison with the previous CdTe detectors. Both the improved charge-collection efficiency and the low leakage current lead to a good energy resolution even at room temperature. Large-area CdTe diode detectors with dimensions of 21.5 × 21.5 mm 2 are now available. By stacking 40 layers of these large-area devices, we have achieved a good energy resolution of several keV (FWHM) for MeV γ-rays.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.