Abstract

We used ROC plots to evaluate the clinical performance of the Guthrie, Wallac, and Isolab assays for newborn phenylketonuria (PKU) screening and assessed the screening discriminatory power of these three assays by the area under the ROC plot, Youden's J index, and the likelihood ratio. The use of these plots not only allows us to pinpoint the exact cutoff value in screening, but also provides a direct comparison of these three different assays in clinical outcome performance. The optimum cutoff for the newborn PKU screening is a blood phenylalanine concentration of 0.30, 0.27, and 0.18 mmol/L for the Guthrie, Wallac, and Isolab assays, respectively. We conclude that the Wallac and Isolab kits, like the Guthrie assay, are suitable for newborn PKU screening.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.